In an era where water quality remains a critical concern for public health, understanding the analytical methods used to detect contaminants is essential. EPA Method 533 is one of two EPA methods approved for measuring per- and polyfluoroalkyl substances (PFAS) in drinking water. PFAS, often referred to as “forever chemicals,” are linked to health effects including certain cancers and thyroid disease, and the parts-per-trillion concentrations at which they matter place heavy demands on laboratory methods. This article covers what Method 533 measures, how it works, how it differs from its companion method, and what utilities and laboratories need to know to use it well.
Method 533 is performed by accredited laboratories rather than in the field, and choosing a laboratory with the right certifications and track record is covered in our guide to selecting a PFAS testing lab.
EPA published Method 533 in late 2019 to extend drinking water PFAS analysis to shorter-chain compounds that earlier methods did not capture well. Shorter-chain PFAS had become increasingly common as manufacturers moved away from long-chain compounds such as PFOA and PFOS, and monitoring needed to keep pace with that shift.
Method 533 quantifies 25 PFAS in drinking water. Its analyte list emphasizes short-chain compounds, beginning with four-carbon acids, and includes perfluorinated carboxylic acids, perfluorinated sulfonic acids, fluorotelomer sulfonates, and several ether-linked PFAS including HFPO-DA (the compound known as GenX). PFOA and PFOS are among its analytes.
Method 533 is best understood alongside its companion. EPA Method 537.1 measures 18 PFAS and has a longer track record; Method 533 measures 25, with stronger coverage of short-chain compounds. The two methods overlap substantially, and together they cover 29 PFAS. That combined list was the basis for nationwide PFAS monitoring under EPA’s fifth Unregulated Contaminant Monitoring Rule, and both methods are approved for compliance monitoring under the federal PFAS drinking water regulation.
The key technical differences:
Both methods sit within a wider family of analytical approaches, from targeted methods for other matrices to non-targeted screening, covered in our overview of PFAS analytical methods.
PFAS are a large family of synthetic compounds — numbering in the thousands — used in industrial applications and consumer products for their resistance to heat, water, and oil. Common sources include firefighting foam, water-repellent textiles, and non-stick coatings. Their resistance to degradation makes them persistent in the environment, and they have been detected in drinking water supplies across much of the United States.
Scientific studies have linked PFAS exposure to developmental effects, immune suppression, elevated cholesterol, and certain cancers. These findings underpinned EPA’s 2024 drinking water regulation establishing enforceable maximum contaminant levels for PFOA and PFOS at 4.0 parts per trillion — concentrations that only methods of the sensitivity of Method 533 and 537.1 can reliably measure.
Sample collection is where PFAS results are most often compromised, and the materials used matter critically. Method 533 samples are collected in polypropylene bottles with polypropylene caps. Fluoropolymer materials such as PTFE must be avoided entirely — fluoropolymers are themselves PFAS-containing materials and are a recognized source of sample contamination.
Samples are preserved with ammonium acetate, which also removes free chlorine that could otherwise affect certain analytes. They must be kept chilled during shipment and stored cold in the laboratory, and extraction must take place within the method’s specified holding time. A field reagent blank — reagent water carried to the sampling site and handled alongside the sample — is required to demonstrate that contamination did not occur during collection. The detailed field practices that protect sample integrity are covered in our guidance on PFAS sampling protocol.
Samples are spiked with isotopically labeled standards, then passed through weak anion exchange solid-phase extraction cartridges that concentrate the PFAS. The extract is eluted, concentrated, and analyzed by liquid chromatography with tandem mass spectrometry (LC-MS/MS). The liquid chromatograph separates the compounds, and the mass spectrometer identifies and quantifies each one by its characteristic mass transitions.
Quality control is integral to Method 533. Required elements include:
Background contamination is a real challenge in PFAS laboratories, since fluoropolymers are common in laboratory equipment. Laboratories performing Method 533 must eliminate PFAS-containing components from their instruments and consumables.
Method 533 is approved for compliance monitoring under the federal PFAS drinking water regulation finalized in April 2024. In May 2026, EPA proposed retaining the PFOA and PFOS limits with an optional compliance extension to 2031, and rescinding the limits for PFHxS, PFNA, HFPO-DA, and the hazard index mixture. As of September 2026 these remain proposals. Because Method 533 measures all of these compounds, it remains relevant whichever way the proposals are resolved, and state programs with their own PFAS standards frequently specify it as well.
For compliance with the federal regulation, either approved method is acceptable. Method 533 is generally preferred where short-chain PFAS are of interest or where the matrix is challenging, because isotope dilution corrects for matrix effects that can bias results. Utilities should confirm which method their state program requires and which their laboratory is certified to perform.
Isotope dilution is the principal source of Method 533’s accuracy advantage. Correcting each result against a labeled analog of the same compound compensates for extraction losses and matrix suppression, which is particularly valuable at the very low concentrations now relevant to compliance.
As PFAS manufacturing has shifted toward shorter-chain compounds, monitoring that captures them has become more important. Short-chain PFAS also break through treatment media earlier than long-chain compounds, so Method 533 data is valuable not only for compliance but for managing treatment system performance.
Contamination during sampling or in the laboratory remains the most common source of unreliable PFAS results. Strict adherence to sampling protocols, appropriate materials, and blank analysis is essential.
Method 533 measures 25 specific compounds. The PFAS family contains thousands of substances, including precursors that can transform into the measured compounds over time or during treatment. Complementary approaches such as the total oxidizable precursor assay estimate the precursor burden that targeted methods cannot see directly.
Method 533 is validated for drinking water. Wastewater, groundwater, soil, and biosolids require different methods designed for those matrices.
More sensitive mass spectrometers, high-resolution instruments capable of non-targeted screening, and total organic fluorine measurements are extending what laboratories can detect. These tools complement rather than replace targeted compliance methods.
Field-deployable screening tools are under development, but for compliance purposes, laboratory LC-MS/MS analysis using approved methods remains the requirement.
EPA Method 533 is a cornerstone of drinking water PFAS monitoring. Its isotope dilution approach and short-chain coverage make it the more robust of the two approved methods for many applications, and its analyte list spans every compound in the federal regulation as it currently stands.
For environmental professionals, the practical priorities are consistent: use a laboratory certified for the method, follow sampling protocols precisely with non-fluoropolymer materials, include field reagent blanks, and understand what targeted analysis does and does not reveal. Staying current on both the method and the evolving regulatory framework will remain essential as PFAS requirements continue to develop.